Description
REL-150-1006-CHCO INTEGRATED POWER DESIGN POWER SUPPLY CHASSIS & COVER 24VDC 6.3A ITE & MEDICAL CERTIFICATION
RoHS?Compliant
Universal 85-264 VAC Input
High Efficiency
Advanced SMT Design
Compact 3.8” x 6.0” x 1.3” Size
2 Year Warranty
Fits 1U Applications
NPUT SPECIFICATIONS |
|
Source Voltage Frequency Range Peak Inrush Current Efficiency Power Factor |
85 – 264 Volts AC 47 – 63 Hz 40A 82% Typ., Full Power, 230V, varies by model 0.95 (Full Power, 230V) |
OUTPUT SPECIFICATIONS |
|
Total Output Power at 50°C – Output Voltage Centering – Output Voltage Adjust Range Load Regulation V1 V2 – – V3 V4 Source Regulation Cross Regulation Output Noise Turn on Overshoot Transient Response Voltage Deviation Recovery Time Load Change Output Overvoltage Protection Output Overpower Protection Hold Up Time Start Up Time |
100W Convection Cooled 150W 300 LFM Forced Air Cooled Output 1: ± 0.5% (All outputs at 50% load) Output 2,3,4: ± 5.0% Output 1: 95 – 105% Output 1: 0.5% (10-100% load change) Output 2: 5.0% (4001-5) 8.0% (2001) 6.0% Output 3: 5.0% Output 4: 5.0% Outputs 1 – 4: 0.5% Outputs 2 – 4: 5.0% Outputs 1 – 4: 1.0% None Outputs 1 – 4 5.0% 500µS 50% to 100% Output 1: 110% to 150% 110-160% rated Pout, cycle on/off, auto recovery 16 mS min., Full Power, 85V Input 5 Seconds, 120V Input |
SAFETY & ELECTROMAGNETIC COMPATIBILITY SPECIFICATIONS |
|
Safety Approvals – – Electrostatic Discharge Radiated Electromagnetic Field EFT/Bursts Surges Conducted Immunity Voltage Interruptions Radiated Emissions Conducted Emissions Harmonic Current Emissions Voltage Fluctuations & Flicker Voltage Dips and Interruptions |
EN 60950-1/A12:2011 EN 60601-1/A2:1995 , EN 60601-1:2006 – EN 61000-4-2, ±8kV Contact Discharge , ±8kV Air Discharge EN 61000-4-3, 80MHz-2.5GHz, 10/m, 80% AM EN 61000-4-4, ±2 kV EN 61000-4-5, ± 1 kV Common Mode ± 2 kV Differential Mode EN 61000-4-6 .15 to 80MHz, 10V, 80% AM EN 61000-4-11, 95% Reduction, 5s EN 55011/22, Class B EN 55011/22, Class B EN 61000-3-2 EN 61000-3-3 EN 61000-4-11, 30% Reduction, 500ms 95% Reduction, 10ms 60% Reduction, 1s (Criteria B) 95% Reductions, 5000ms |
ENVIRONMENTAL SPECIFICATIONS |
|
Ambient Operating Temperature Range Ambient Storage Temp. Range Temperature Coefficient |
0° C to + 70° C Derating: See Power Rating Chart – 40° C to + 85° C Outputs 1 – 4: 0.02% /°C |
GENERAL SPECIFICATIONS |
|
Means of Protection Primary to Secondary Primary to Ground Secondary to Ground Dielectric Strength(15) Reinforced Insulation Basic Insulation Operational Insulation Leakage Current Earth Leakage Touch Current – Power Fail Signal – Remote On/Off (optional) Remote Sense Mean-Time Between Failures Weight |
– 2MOPP (Means of Patient Protection) 1MOPP (Means of Patient Protection) Operational Insulation(Consult factory for 1MOOP or 1MOPP) (See Datasheet) 5656 VDC, Primary to Secondary, 1 Sec. 2545 VDC, Primary to Ground, 1 Sec. 707 VDC, Secondary to Ground, 1 Sec. – less than 300uA NC, less than 1000uA SFC less than 100uA NC, less than 500uA SFC – Logic low with input power failure 10 mS minimum prior to Output 1 dropping 1% Contact closure shuts off all outputs 250mV compensation of output cable losses 100,000 Hours min., MIL-HDBK-217F, 25° C, GB 1.15 Lbs. Open Frame/ 1.82 Lbs. Chassis and Cover |
Manufacturer Part Number: REL-150-1006-CHCO
Manufacturer: Electronic Components